• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING Back to Search Results
Model Number 442021
Device Problem Contamination of Device Ingredient or Reagent (2901)
Patient Problem Bacterial Infection (1735)
Event Date 05/03/2022
Event Type  Injury  
Event Description
It was reported that customer had several instances of identified organisms they suspect to be contamination from bd bactec¿ lytic/10 anaerobic/f culture vials (plastic)- lot #2040802.Three patients affected by decision to administer antibiotics.
 
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that customer had several instances of identified organisms they suspect to be contamination from bd bactec¿ lytic/10 anaerobic/f culture vials (plastic)- lot #2040802.Three patients affected by decision to administer antibiotics.
 
Manufacturer Narrative
H6: investigation summary : catalog 442021; batch no.2040802.Customer reported a contamination issue while using bactec product.Returned good samples were received.Bd was unable to reproduce the customer¿s experience with bactec product.Retention/returned samples were visually inspected, tested for viable contamination by sub-culturing on tsa, chocolate, sabouraud and schaedler agars plates, voltage output and gram stain.All results were satisfactory.Batch history record review did not identify any evidence for which the customer submitted the complaint.A complaint history review was conducted and only the current complaint was found relating to the incident lot number and the ¿as reported¿ defect code.Complaint is unconfirmed based on retention/returned samples and batch history record review.Product insert warnings and precautions sections states that prior to use, each vial should be examined for evidence of contamination such as cloudiness, bulging or depresses septum, or leakage.Vials showing evidence of contamination should not be used.Also prior to use, the user should examine the vial for evidence of damage or deterioration.Vials displaying turbidity, contamination, or discoloration (darkening) should not be used.No corrective actions were required.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigational process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)
Type of Device
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14889733
MDR Text Key295151843
Report Number2647876-2022-00167
Device Sequence Number1
Product Code MDB
UDI-Device Identifier00382904420215
UDI-Public00382904420215
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123903
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/30/2022
Device Model Number442021
Device Catalogue Number442021
Device Lot Number2040802
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/06/2022
Initial Date FDA Received07/01/2022
Supplement Dates Manufacturer Received07/06/2022
Supplement Dates FDA Received07/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-