PHILIPS NORTH AMERICA LLC HEARTSTART MRX MONITOR/DEFIB; DEFIBRILLATOR Back to Search Results
Model Number M3535A
Device Problems Device Alarm System (1012); Failure of Device to Self-Test (2937)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/29/2022
Event Type  malfunction  
Event Description
It was reported to philips that the device responds with an error.There was reportedly no patient involvement.
 
Manufacturer Narrative
The manufacturer's authorized field service engineer (fse) evaluated the device and confirmed the reported problem.Upon conclusion of the evaluation, it was determined that this was a malfunction caused by configuration settings, the settings were reset, and the device was returned to full functionality.The device remains at the customer site and no further evaluation is warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTSTART MRX MONITOR/DEFIB
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
daniel derochers
22100 bothell everett highway
bothell, WA 98021
9095703538
MDR Report Key14895795
MDR Text Key295339000
Report Number3030677-2022-03317
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00884838000018
UDI-Public00884838000018
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K031187
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3535A
Device Catalogue NumberM3535A
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/29/2022
Initial Date FDA Received07/01/2022
Supplement Dates Manufacturer Received06/29/2022
Supplement Dates FDA Received10/03/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-