Additional information added to h3, h6 and h10 h10: the device was evaluated on site by a qualified technician who confirmed the reported problem.During evaluation, it was noted that there was a communication error and the hardware was broken.It was found that the machine casing, slave pump, back pressure connector , inlet pressure connector, heparin pump and lord motor were broken.These were replaced to resolve the issues.The cause was most likely related to uneven ground, only one nurse transferring the equipment and the equipment in an old version of the base.The qualified technician recommended that the hospital; arrange for more than two people to transfer equipment.It was also recommended that the hospital replace the base to reduce the risk of falling.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
|