This follow-up report is being submitted to relay additional and corrected information.The following sections were updated: b4, d9, g3, g6, h2, h3, h6, h10.The following sections were corrected: d4 visual evaluation of the returned product confirmed the sterile pouch was likely punctured by the distal tip of the stem, resulting in the pouch tearing.The device history records were reviewed and no discrepancies were identified.The condition of the device when it left zimmer biomet is conforming to specification and therefore the root cause of the reported event is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|