Model Number N/A |
Device Problem
Device Contaminated During Manufacture or Shipping (2969)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: japan.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that there was debris in the sterile package.No adverse events have been reported as a result of the malfunction and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Evaluation of the returned product/photographs provided confirmed foreign debris is present inside the sterile packaging, and the sterile packaging remains sealed.The reported event is confirmed.Device history record was reviewed and no discrepancies related to the reported event were found.The likely condition of the device when it left zimmer biomet is nonconforming to specification.The root cause of the reported event is the operator not following the provided work instructions during manufacturing.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional event information to report at this time.
|
|