RESPIRONICS CALIFORNIA, LLC RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 VENT, JAPAN OPT: CFLEX,AVAPS,AT+ |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/20/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).
|
|
Event Description
|
The customer reported that the waveform in the flow and the volume were not displayed on the device.The device was in use at the time of the event.The ventilator was swapped for an alternate ventilator.No patient or user harm reported.The customer reported that the day after the device was removed from service, an operation check was performed on the device, but the reported phenomenon was not observed.It was then noted by the philips sales representative, that after collecting the device, it was heard that the flow had been provided but the waveform(s) had not been displayed on the display.Further information has been requested.If additional information becomes available at a later date, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
H10: the reported phenomenon was not duplicated.Functional testing was performed for inspections but the result was that any anomaly was not found.
|
|