The 95-3000 t15 driver was not returned for investigation, nor were any photos of the complaint unit received.Review of the device history record concluded the instrument is approximately 4.5 years old.The root cause cannot be determined as the driver was not provided for evaluation.Additionally, no photos of the complaint unit were received.Based on prior complaints related to driver fractures, analyses suggest the tip failure is the result of field damage or age/wear, including but not limited to repeat use, reprocessing, or excessive torsional force.The respective driver is a reusable instrument that is subject to handling and multiple sterilization cycles after distribution into the field.Wear and damage can occur over time.
|