U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER; GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number N/A
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/09/2021
Event Type  Injury  
Event Description
It was reported that an ultrathane mac-loc locking loop multipurpose drainage catheter was discovered to have migrated out of the intended cavity nine days after placement.The patient underwent drainage catheter placement on (b)(6) 2021 for abscess drainage in the pericolonic lower abdomen.Computed tomography was utilized for visual guidance and dilation was performed prior to catheter placement.On (b)(6) 2021, nine days after placement, a ct identified that the catheter had retracted outside of the intended cavity.Complete fluid evacuation was unable to be achieved with the device.As a result, the device was removed and replaced on (b)(6) 2021.The patient did not experience any additional adverse effects prior to exiting the study.
 
Manufacturer Narrative
Occupation: unknown.Pma/510(k) #: exempt.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Investigation ¿ evaluation: it was reported by cook research inc.Via (b)(6) medical center (usa) that an ultrathane mac-loc locking loop multipurpose drainage catheter (rpn: ult8.5-38-25-p-6s-clm-rh; lot#: 13804335) migrated nine days after placement.On (b)(6) 2021, the patient underwent a study procedure (mdr-1919) to place the catheter for abscess drainage in the pericolonic lower abdomen.Dilation was performed prior to catheter insertion, then the catheter was placed using computed tomography (ct) for visual guidance.On (b)(6) 2021, a ct identified that the catheter had retracted outside of the intended cavity.Complete fluid evacuation was unable to be achieved with the device.As a result, the patient required surgical intervention to replace the device on (b)(6) 2021.The patient did not experience any additional adverse effects prior to exiting the study.No other adverse events were reported due to this occurrence.Reviews of the complaint history, device history record (dhr), quality control procedures, manufacturing instructions and instructions for use (ifu) of the device, were conducted during the investigation.The complaint device was not returned to the manufacturer for evaluation; therefore, no physical examinations could be conducted.However, a document-based investigation evaluation was performed.A review of the device master record (dmr) concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.A review of the dhr for the reported complaint device lot revealed no relevant non-conformances.A database search did not identify any other events associated with the reported device lot.Based on the available information, cook has concluded that the device was manufactured to specification and that there is no evidence suggesting nonconforming product exists either in house or in field.Cook also reviewed product labeling.The instructions for use (ifu) [ t_multi_rev5, multipurpose drainage catheter] states the following."how supplied: upon removal from package, inspect the product to ensure no damage has occurred.Intended use: multiple drainage catheters are intended for percutaneous drainage applications (e.G nephrostomy, biliary and abscess), either by direct stick or seldinger access technique.Warnings: if a catheter has become malposition or if drainage ceases, the catheter should be promptly exchanged or removed.References: these instructions for use are based on experience from physicians and (or) their published literature.Refer to your local cook sales representative for information on available literature.¿ based on the available information, no product returned, and the results of the investigation, a root cause was unable to be established.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRATHANE MAC-LOC LOCKING LOOP MULTIPURPOSE DRAINAGE CATHETER
Type of Device
GBO CATHETER, NEPHROSTOMY, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC.
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14947619
MDR Text Key295439198
Report Number1820334-2022-01167
Device Sequence Number1
Product Code GBO
UDI-Device Identifier00827002095014
UDI-Public(01)00827002095014(17)240305(10)13804335
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Study,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/05/2024
Device Model NumberN/A
Device Catalogue NumberULT8.5-38-25-P-6S-CLM-RH
Device Lot Number13804335
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/22/2022
Initial Date FDA Received07/07/2022
Supplement Dates Manufacturer Received11/29/2022
Supplement Dates FDA Received12/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient Weight152 KG
-
-