OLYMPUS MEDICAL SYSTEMS CORP. ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-PRO
Device Problem Restricted Flow rate (1248)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2022
Event Type  malfunction  
Manufacturer Narrative
During troubleshoot the customer removed the water supply hone and cleaned the filter at the incoming water port.The filter was cleaned as described in the instructions for use (7.13).The filter was cleaned but the customer indicated the filter still has residue.Replacement parts have been ordered by the olympus field service engineer.The investigation is ongoing; therefore, the root cause of the reported issue cannot be determined at this time.However, if additional information becomes available this medical device report will be supplemented accordingly.
 
Event Description
Olympus technical support engineer was informed via service inquiry that the customer endoscope reprocessor has reduced flow of water supply due to excessive debris residue on the mesh filter in the water supply hose connector which has resulted in, ¿increased cycle times.¿ the cycle time increased to 33+ minutes.No death, injury or harm had been reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed the reduced water flow into the reprocessing basin.A definitive root cause for this issue was not established.However, it is probable that the issue occurred because of a clogged filter.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14953851
MDR Text Key304196936
Report Number8010047-2022-11557
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/09/2022
Initial Date FDA Received07/08/2022
Supplement Dates Manufacturer Received07/26/2022
Supplement Dates FDA Received08/24/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-