ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER HAND PIECE Back to Search Results
Catalog Number 8065750469
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Conjunctivitis (1784); Eye Infections (4466); Eye Pain (4467); Toxic Anterior Segment Syndrome (TASS) (4469); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 04/08/2022
Event Type  Injury  
Event Description
A physician reported that a patient experienced toxic anterior segment syndrome, slight pain, tyndall, conjunctival inflammation, aqueous cells, conjunctival infection when an ophthalmic viscoelastic was used.Patient was given with steroids, antibiotics and non-steroidal anti-inflammatory medications.Symptoms were resolved.No further information expected as customer was unwilling to provide.Additional information received clarifying that the reporter suspects ophthalmic phacoemulsification handpiece for the events.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The handpiece was not returned for evaluation.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.A potentially relevant complaint was found and reviewed as part of this investigation.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.The root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HAND PIECE
Type of Device
UNK
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key14959661
MDR Text Key295520778
Report Number2028159-2022-00900
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065750469
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/28/2022
Initial Date FDA Received07/08/2022
Supplement Dates Manufacturer Received07/14/2022
11/16/2022
Supplement Dates FDA Received08/11/2022
12/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/14/2008
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ACUNEX VARIO; CENTURION VISION SYSTEM; OZIL TORSIONAL HANDPIECE; VISCOSURGICAL DEVICE
Patient Outcome(s) Required Intervention; Other;
Patient Age64 YR
Patient SexFemale
-
-