It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip could not be deployed despite several attempts.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
Block h6: medical device problem code a15 captures the reportable event of clip would not release from the catheter.Block h10: the returned resolution 360 clip device was analyzed, and a visual evaluation noted that the device was not returned with over sheath.The coil is unraveled next to the bushing.The device did have the clip assembly attached.No activations were found on the clip.The reported event of the clip the wound was clipped, but the clip could not deploy after several attempts, could not be confirmed.Investigation found, the device returned with the clip assembly still attached and it was able to open and close its arms.As part of the analysis, the most distal section was inspected and not discernible failures were observed.The functional inspection was performed using the tortuous fixture as per procedure 91158713, clip deployment test procedure, bsc, ver.M, and it was able to deploy without problems.In addition, the device returned without the over sheath, just with the blue grip attached, but, since there is not enough evidence to clarify the absence of this component, it will be classified as "cause not established".Also, the coil is unraveled next to the bushing, this failure can be caused due to operational factors such as the over sheath removal or insertion/removal from the scope.Therefore, this failure will be classified as "adverse event related to procedure".The investigation findings, and all the information available lead to no problem detected.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.Based on the information provided, there is no evidence that the device was used in a manner inconsistent with the labelled indications 51259680, dfu, mb resolution clip, bsc, ver a.No further review is required.
|
It was reported to boston scientific corporation that a resolution clip device was used in the colon during an endoscopic mucosal resection (emr) procedure on june 22, 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip could not be deployed despite several attempts.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|