• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RAYNHAM CODMAN DISPOS PERFORATOR; DISPOSABLE PERFORATORS Back to Search Results
Model Number 261221
Device Problem Dull, Blunt (2407)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
(b)(4).A facility reported a disposable perforator (id 261221) perforated poorly through bone and it was very dull.The surgeon had to stopped mid burr hole to get another perforator.The drill used is pneumatic codman drill.The perforator clicked in placed in the drill, and the recommended spring tests were performed between each burr hole.They also stated that, it took over 5 minutes to drill each burr hole.The procedure was completed with another product available.
 
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
N/a.
 
Manufacturer Narrative
The disposable perforator (id 261221) was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.However, a probable root cause for the reported complaint includes insufficient sharpness specification or user misuse.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CODMAN DISPOS PERFORATOR
Type of Device
DISPOSABLE PERFORATORS
Manufacturer (Section D)
RAYNHAM
325 paramount drive
325 paramount drive
raynham MA
Manufacturer (Section G)
RAYNHAM
325 paramount drive
raynham MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15001127
MDR Text Key304694287
Report Number3014334038-2022-00158
Device Sequence Number1
Product Code HBF
UDI-Device Identifier10381780513599
UDI-Public10381780513599
Combination Product (y/n)N
PMA/PMN Number
K791101
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number261221
Device Catalogue Number261221
Device Lot Number5988978
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/27/2022
Initial Date FDA Received07/13/2022
Supplement Dates Manufacturer Received08/17/2022
Supplement Dates FDA Received08/31/2022
Date Device Manufactured01/05/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-