• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO AUTOLOG WASH KIT; APPARATUS, AUTOTRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO AUTOLOG WASH KIT; APPARATUS, AUTOTRANSFUSION Back to Search Results
Model Number ATL2001
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2022
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during use of an autolog wash kit, it was reported there was an issue with the collection bag in the kit.The customer was unable to get the washed blood out of the collection bag after they successfully removed one washed bowl, to a transfer bag to the patient.When they tried to remove the next bowl of washed blood from the collection bag, they could not do it.They tried to move it to a transfer bag, tried spiking directly into the bag with iv tubing, and attempted to use a pressure bag, but blood would not flow out of the bag.This issue occurred with both ports in of the bag.In the end the customer ended up discarding the washed blood since they were unable to remove it from the bag.There was no patient impact associated with this event.Medtronic received additional information stating that the amount of patient blood loss as a result of discarding the washed blood was 200 ml.The customer stated that it is unknown if the patient required a transfusion because of the discarded blood.The customer stated that there was no impact or injury to the patient.There were no blockage or clots observed in the tubing or the bags.There was no damage or other issues noted with the tubing or the bags.Medtronic received additional information that the customer did not open another wash kit to complete the case after the issue occurred or replace the collection bag as the issue was observed at the end of the case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOLOG WASH KIT
Type of Device
APPARATUS, AUTOTRANSFUSION
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15011641
MDR Text Key301658558
Report Number9612164-2022-02656
Device Sequence Number1
Product Code CAC
UDI-Device Identifier00613994520456
UDI-Public00613994520456
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093535
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/17/2024
Device Model NumberATL2001
Device Catalogue NumberATL2001
Device Lot Number223942455
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/15/2022
Initial Date FDA Received07/14/2022
Date Device Manufactured03/17/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-