• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND METAGLENE; SHOULDER METAGLENE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND METAGLENE; SHOULDER METAGLENE Back to Search Results
Model Number 130760000
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Foreign Body Reaction (1868); Discomfort (2330)
Event Date 06/27/2022
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Surgeon reported a revision of left glenosphere.The patient experienced discomfort since 3-4 weeks.The screw pin including the plate have detached from the shell.Metallosis in the joint.Doi: (b)(6) 2021.Dor: (b)(6) 2022.Left shoulder.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Additional information received indicated that the reason for surgery was humeral head comminuted fracture.No surgical time extension occurred during the revision on (b)(6) 2022.The central screw pin was broken but could be removed without problems.The metaglene locking screws are tight.Thus, only revision of the glenosphere and inlay change.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND METAGLENE
Type of Device
SHOULDER METAGLENE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15013367
MDR Text Key295866630
Report Number1818910-2022-13027
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027737
UDI-Public10603295027737
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date 08/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130760000
Device Catalogue Number130760000
Device Lot Number5374336
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/27/2022
Initial Date FDA Received07/14/2022
Supplement Dates Manufacturer Received07/27/2022
07/27/2022
08/30/2022
Supplement Dates FDA Received08/01/2022
08/08/2022
08/30/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/18/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DXTEND GLENOSPHERE STD D42MM; DXTEND HIGHMO PE CUP D42 +6MM; DXTEND HIGHMO PE CUP D42 +9MM; DXTEND MBLOC HUM EPI 1 D10 STD; DXTEND SCREW LOCK D4.5X24MM; DXTEND SCREW LOCK D4.5X36MM; DXTEND SCREW NO LOCK D4.5X24MM; DXTEND SCREW NO LOCK D4.5X30MM; UNKNOWN SHOULDER GLENOSPHERE
Patient Outcome(s) Required Intervention;
Patient Age72 YR
Patient Weight100 KG
-
-