Model Number MCH-1000 (I)
Device Problem Blocked Connection (2888)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/02/2020
Event Type  malfunction  
Manufacturer Narrative
The manufacturer is reporting the following complaint after a voluntary review of all complaints (reportable or not) since 2016.This report is being filed now, after being scrutinized under a newly revised risk matrix, recently adopted after inspection.Cardioquip was notified by the customer that there was patient involvement during the malfunction, although no information identifying the patient was provided.Upon initial investigation, cardioquip determined that the unit's water level sensor was improperly placed which allowed the system's main pump to operate without water in the line.The water level sensor was relocated to the correct location and all damaged components were replaced.Following the repairs, the device passed inspection and is fully operational.
 
Event Description
Customer reports bruning smell coming from device while priming for a case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOQUIP MODULAR COOLER HEATER
Type of Device
CARDIOPULMONARY BYPASS DEVICE
Manufacturer (Section D)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer (Section G)
CARDIOQUIP, LLC
8422 calibration ct.
college station TX 77845
Manufacturer Contact
charley ford
8422 calibration ct.
college station, TX 77845
9796910202
MDR Report Key15013494
MDR Text Key304709192
Report Number3007899424-2022-00063
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102147
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Inspection
Type of Report Initial
Report Date 05/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberMCH-1000 (I)
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/18/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-