HOYA CORPORATION PENTAX TOKYO OFFICE PENTAX; VIDEO BRONCHOSCOPE 1.2C 3.8TP 600L Back to Search Results
Model Number EB-1170K
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2022
Event Type  malfunction  
Manufacturer Narrative
We checked the returned unit and confirmed that the operation channel stuck accessories and objects.Based on the result, we concluded that it was caused due to the insufficient reprocessing at the facility on the operation channel.In addition, we confirmed that the light guide fiber bundle (lcb) broken, the insertion flexible tube (ift) buckled, the light guide cable buckled, the remote control buttons cracked, and the distal body worn out; however, they are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report "hr-rpt-0588(channel)" and/or the risk analysis results, it was evaluated to submit mdr.
 
Event Description
The time of event is unknown.There was no report of patient harm.Operation/suction channel clogged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTAX
Type of Device
VIDEO BRONCHOSCOPE 1.2C 3.8TP 600L
Manufacturer (Section D)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA  196-0012
Manufacturer (Section G)
HOYA CORPORATION PENTAX TOKYO OFFICE
tsutsujigaoka 1-1-110
akishima-shi, tokyo 196-0 012
JA   196-0012
Manufacturer Contact
william goeller
3 paragon drive
montvale, NJ 07645
MDR Report Key15025291
MDR Text Key300419821
Report Number9610877-2022-55619
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131028
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEB-1170K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2022
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/22/2022
Initial Date FDA Received07/15/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-