Model Number N/A
Device Problem Off-Label Use (1494)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/20/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Source foreign: italy.Investigation of this incident is currently ongoing.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
 
Event Description
It was reported that expired product has been implanted.No known impact to the patient.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.Sterile cert reviewed and confirms product was irradiated in conformance with technical agreement.Medical records were not provided.The likely condition of the device when it left zimmer biomet is conforming to specification.The definitive root cause of the reported event cannot be determined with the information provided, however user error is likely to have contributed to the reported event.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOLOX DELTA HEAD, 28MM +3.5MM NECK, 12/14 TAPER
Type of Device
HIP PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15025898
MDR Text Key295954885
Report Number3002806535-2022-00307
Device Sequence Number1
Product Code LZO
UDI-Device Identifier05019279116881
UDI-Public(01)05019279116881(17)220619(11)2732825
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date06/19/2022
Device Model NumberN/A
Device Catalogue Number650-0832
Device Lot Number2732825
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/20/2022
Initial Date FDA Received07/15/2022
Supplement Dates Manufacturer Received07/29/2022
Supplement Dates FDA Received08/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/03/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
-
-