Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Weight Changes (2607); Swelling/ Edema (4577)
Event Date 03/08/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3007963827-2022-00184, 3007963827-2022-00185, 0002648920-2022-00162, 0001822565-2022-02100, 0001822565-2022-02101.
 
Event Description
It was reported a patient was converted from a right medial uncompartmentalized knee to a right total knee.Subsequently, the patient is experiencing ongoing pain, swelling, radiolucency at the medial tibial plateau, and possible instability.Anti-inflammatory gel and steroids were prescribed to relieve the symptoms.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: swelling, possible instability, full rom, prescribed medication for swelling, x-rays show radiolucency in the medical tibial plateau.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIA CEMENTED 5 DEGREE STEMMED RIGHT SIZE E
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15026360
MDR Text Key295954171
Report Number0001822565-2022-02099
Device Sequence Number1
Product Code MBH
UDI-Device Identifier00889024471061
UDI-Public(01)00889024471061(17)280930(10)64159725
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172524
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42532007102
Device Lot Number64159725
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/24/2022
Initial Date FDA Received07/15/2022
Supplement Dates Manufacturer Received09/29/2022
Supplement Dates FDA Received10/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/05/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
Patient Weight123 KG
-
-