MAUDE Adverse Event Report: KARL STORZ ENDOVISION, INC. CMOS VIDEO CYSTO-URETHROSCOPE C-VIEW REV DEF; FLEXIBLE VIDEO CYSTO SCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ ENDOVISION, INC. CMOS VIDEO CYSTO-URETHROSCOPE C-VIEW REV DEF; FLEXIBLE VIDEO CYSTO SCOPE Back to Search Results
Model Number 11272VUEK
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2022
Event Type  malfunction  
Manufacturer Narrative
Product has not been returned by the facility.
 
Event Description
During a procedure, image on screen would black out while hitting the cautery button.No patient injury was reported.
 
Manufacturer Narrative
Per evaluation findings: image interruption due to esu (electrosurgical unit) failure, ccu cable.Over bend on shaft 60mm from shaft adaptor.Dents on distal tip.Image interruption due to esu (electrosurgical unit) failure, ccu cable.This failure mode was confirmed and has been determined to be due to the durability of the power cable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CMOS VIDEO CYSTO-URETHROSCOPE C-VIEW REV DEF
Type of Device
FLEXIBLE VIDEO CYSTO SCOPE
Manufacturer (Section D)
KARL STORZ ENDOVISION, INC.
91 carpenter hill road
charlton MA 01507
Manufacturer (Section G)
KARL STORZ ENDOVISION
91 carpenter hill road
charlton MA 01507
Manufacturer Contact
anja fair
2151 e. grand avenue
el segundo, CA 90245-5017
4242188738
MDR Report Key15027541
MDR Text Key295972624
Report Number1221826-2022-00120
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K202957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number11272VUEK
Device Catalogue Number11272VUEK
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/13/2022
Initial Date FDA Received07/15/2022
Supplement Dates Manufacturer Received06/13/2022
Supplement Dates FDA Received12/06/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-