• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ULTRAFLEX TRACHEOBRONCHIAL; PROSTHESIS, TRACHEAL, EXPANDABLE Back to Search Results
Model Number M00564840
Device Problem Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).The complainant indicated that the ultraflex tracheobronchial stent was contaminated and will not be returned for evaluation; therefore, a device analysis of could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation on june 24, 2022 that an ultraflex tracheobronchial covered distal release stent was to be implanted in the upper part of the lungs to treat a 3cm lung cancer during a bronchoscopy procedure performed on (b)(6), 2022.The patient's anatomy was tight and was dilated using a cre pulmonary balloon dilatation catheter prior to stent placement procedure.During the procedure, the stent was deployed; however, the stent was unable to expand.The stent was removed from the patient with rescue rat-tooth forceps and another ultraflex tracheobronchial stent of different size was used to complete the procedure.There were no patient complications as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.Note: a photo of the device provided by the complainant outside the patient shows the stent was not fully expanded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAFLEX TRACHEOBRONCHIAL
Type of Device
PROSTHESIS, TRACHEAL, EXPANDABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15032658
MDR Text Key296009972
Report Number3005099803-2022-03835
Device Sequence Number1
Product Code JCT
Combination Product (y/n)N
Reporter Country CodeLE
PMA/PMN Number
K012883
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/04/2023
Device Model NumberM00564840
Device Catalogue Number6484
Device Lot Number0028325611
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/24/2022
Initial Date FDA Received07/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age68 YR
Patient SexMale
-
-