• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. VIDEOSCOPE "CYF-V2", UK VERSION; FLEXIBLE VIDEO CYSTONEPHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. VIDEOSCOPE "CYF-V2", UK VERSION; FLEXIBLE VIDEO CYSTONEPHROSCOPE Back to Search Results
Model Number CYF-V2
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2022
Event Type  malfunction  
Event Description
The customer reported that the bending rubber was torn at the tip, and metal coils were visible where the bending rubber was torn/missing.The issue was observed during reprocessing.There was no patient or user injury reported due to the event.
 
Manufacturer Narrative
The subject device was returned to an olympus service center for evaluation.During inspection and testing, service confirmed the customer¿s reported issue and found there was a leak due to a cut in the bending section cover (a-rubber).The metal braiding beneath the rubber covering was exposed, but it was not protruding.In addition, the adhesive on the bending section cover was cracked.The video connector case and video connector plug were cracked.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to physical stress applied to the device.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEOSCOPE "CYF-V2", UK VERSION
Type of Device
FLEXIBLE VIDEO CYSTONEPHROSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key15032768
MDR Text Key301996085
Report Number8010047-2022-12041
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170339431
UDI-Public04953170339431
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133538
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/15/2022
Initial Date FDA Received07/15/2022
Supplement Dates Manufacturer Received07/29/2022
Supplement Dates FDA Received08/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/06/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-