• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXALT MODEL D SINGLE-USE DUODENOSCOPE; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXALT MODEL D SINGLE-USE DUODENOSCOPE; DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number M00542421
Device Problems Break (1069); Optical Problem (3001); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/17/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complainant indicated that the device is not available for return; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an exalt model d single-use duodenoscope was used during an endoscopic retrograde cholangiopancreatography (ercp) performed on (b)(6) 2022 for treatment of biliary obstruction.During the procedure, the physician was attempting to cannulate the papilla for about 20 minutes when visualization inside the patient was lost and the scope error screen appeared.It was also noted that the exalt model d single-use duodenoscope movements were not responding correctly.The physician then removed the exalt model d single-use duodenoscope from the patient and observed that the distal tip of the scope had partially detached from the insertion tube.This procedure was completed with a second exalt model d single-use duodenoscope.There were no patient complication reported as result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXALT MODEL D SINGLE-USE DUODENOSCOPE
Type of Device
DUODENOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15042400
MDR Text Key304615251
Report Number3005099803-2022-03903
Device Sequence Number1
Product Code FDT
UDI-Device Identifier08714729993605
UDI-Public08714729993605
Combination Product (y/n)N
PMA/PMN Number
K193202
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/16/2024
Device Model NumberM00542421
Device Catalogue Number42421
Device Lot Number0028883680
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/20/2022
Initial Date FDA Received07/18/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/16/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-