• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CONSTELLATION VISION SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065753041
Device Problem Failure to Fire (2610)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/15/2022
Event Type  malfunction  
Event Description
A customer reported that an ophthalmic laser turning off during shooting.The surgery was completed successfully.The procedure details and patient impact were not reported.Additional information has been requested, but none received to date.
 
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
The company representative confirmed and replicated the reported event.The footswitch was subsequently replaced to address the issue.The system was tested and found to meet product specifications.A non-conformance-based review of the batch/lot/serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this lot/batch/serial number was performed.No similar complaints were reported for the product lot/batch/serial under investigation.The root cause of the reported event is attributed to nonconforming footswitch.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15043548
MDR Text Key296322547
Report Number2028159-2022-00956
Device Sequence Number1
Product Code HQC
UDI-Device Identifier00380657530410
UDI-Public00380657530410
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065753041
Device Lot Number13P7W4
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/20/2022
Initial Date FDA Received07/18/2022
Supplement Dates Manufacturer Received09/14/2022
Supplement Dates FDA Received10/10/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-