• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR; BIPOLAR SINGLE USE CUTTING LOOP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG CUTTING LOOP, BIPOLAR; BIPOLAR SINGLE USE CUTTING LOOP Back to Search Results
Model Number 27040GP130-S
Device Problem Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2022
Event Type  malfunction  
Event Description
During a procedure, the loop completely came off the tip of the electrode.There was no reported harm to the patient and the procedure was completed.
 
Manufacturer Narrative
Product has not been returned by facility, therefore product has not been evaluated.
 
Manufacturer Narrative
Corrected lot code and added manufactured date.Added the evaluation findings below: the distal cutting loop is dislodged from housing shaft.The distal plastic on the loop and the insulation has melted.Black residue is seen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUTTING LOOP, BIPOLAR
Type of Device
BIPOLAR SINGLE USE CUTTING LOOP
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen, gm
Manufacturer (Section G)
KARL STORZ SE & CO.KG
dr.-karl-storz -strasse 34
78532
tuttlingen, gm
Manufacturer Contact
anja fair
2151 e. grand avenue
el segundo, CA 90245
4242188247
MDR Report Key15053012
MDR Text Key296852535
Report Number9610617-2022-00125
Device Sequence Number1
Product Code FAS
UDI-Device Identifier04048551393864
UDI-Public4048551393864
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number27040GP130-S
Device Catalogue Number27040GP130-S
Device Lot NumberOP02
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/13/2022
Initial Date FDA Received07/19/2022
Supplement Dates Manufacturer Received06/13/2022
Supplement Dates FDA Received08/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-