• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INVAMEX POWER LIFT W/LOW BASE-PLUS 9153633519; LIFT, PATIENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INVAMEX POWER LIFT W/LOW BASE-PLUS 9153633519; LIFT, PATIENT, AC-POWERED Back to Search Results
Model Number NA:RPL450-1
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The bolt in the hanger bar of the patient lift came out resulting in the hanger bar and user in the sling falling in the middle of a transfer.The user was roughly 3 inches above the arms of a chair when they fell.The user was transported to a hospital for evaluation but had no apparent injuries.The device was taken out of use.
 
Manufacturer Narrative
Multiple attempts were made to obtain additional information regarding this incident, without success.The device was over 13 years old at the time of this incident.The user¿s weight was within the weight capacity of 450 pounds listed in the manual.This incident has been previously investigated via a corrective action.Contributing factors including inadequate joint design were found to be the most likely cause of this failure.Corrections are actively being implemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER LIFT W/LOW BASE-PLUS 9153633519
Type of Device
LIFT, PATIENT, AC-POWERED
Manufacturer (Section D)
INVAMEX
102 parque industrial manimex
reynosa, tamaulipas 88780
MX  88780
Manufacturer (Section G)
INVAMEX
102 parque industrial manimex
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
jason fiest
one invacare way
elyria, OH 44035
8003336900
MDR Report Key15074139
MDR Text Key298587071
Report Number9616091-2022-00012
Device Sequence Number1
Product Code FNG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNA:RPL450-1
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/22/2022
Initial Date FDA Received07/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Weight70 KG
-
-