Updated fields: b4, d9(device available for eval), g3, g6, h2, h3, h4, h6(investigation type, investigation findings, component codes & investigation conclusions), h10, h11 corrected fields: b5, d5, e1(initial reporter & event site email), e2, e3, e4, g2, h6(problem code).The getinge field service engineer (fse) that encountered the issue, rectified by replacing the power supply.The unit was tested and found pressure and vacuum were better and is working but intermittently.It is showing leak in iabp circuit message.Drive manifold assy need to be cross checked.The fse replaced the drive manifold from company stock to hospital towards machine under cmc.Also replaced safety disk from hospital stock.The machine working fine and the issue was resolved.The fse completed all safety, functionality, and calibration checks and all tests passed to factory specifications.The iabp was then released to the customer and cleared for clinical service.
|