• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. CPCS COCR PRIM SO 12/14 SZ 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. CPCS COCR PRIM SO 12/14 SZ 3; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 71312363
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Bone Fracture(s) (1870); Hip Fracture (2349)
Event Date 12/01/2011
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that after a left thr performed on (b)(6) 2008, the patient experienced a femoral periprosthetic fracture on the cpcs cocr prim so 12/14 sz 3 implant area.The adverse event was treated via revision surgery on (b)(6) 2011 and the cpcs cocr prim so 12/14 sz 3, the cocr 12/14 fem head 28 +0 and the cpcs dist cent sz 14mm were explanted.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent either a revision or re-revision surgery where a polarcup shell was part of the hip construct.As such, no further information will be available.
 
Manufacturer Narrative
Section h10: the device was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, this reported event was derived from the national joint registry uk summary report for polarcup as part of a retrospective data collection.As such, no further information will be available.Without clinically relevant patient-specific supporting documentation, a thorough medical investigation could not be performed.The root cause and/or patient outcome beyond that which was reported could not be confirmed nor concluded; therefore, no further medical assessment is warranted at this time.Should clinical documentation become available in the future, a thorough medical assessment may be rendered at that time.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for total hip systems revealed that patient should be warned of surgical risks, and made aware of possible adverse effects has been identified in the warnings and precautions/preoperative section.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or potential causes that could contribute to the reported event include excessive forces applied to implant and surgical technique.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CPCS COCR PRIM SO 12/14 SZ 3
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15076727
MDR Text Key296326593
Report Number1020279-2022-03400
Device Sequence Number1
Product Code JDI
UDI-Device Identifier03596010462022
UDI-Public03596010462022
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K823727
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71312363
Device Catalogue Number71312363
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/30/2022
Initial Date FDA Received07/21/2022
Supplement Dates Manufacturer Received08/04/2022
Supplement Dates FDA Received08/30/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
71302800: COCR 12/14 FEM HEAD 28 +0; 71312414: CPCS DIST CENT SZ 14MM; 75008539: POLARCUP LINER 53/28; 75008720: POLARCUP SHELL HA COATED 53 NON CEM
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age78 YR
Patient SexFemale
-
-