• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP SHELL HA COATED 51 NON CEM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG POLARCUP SHELL HA COATED 51 NON CEM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 75008719
Device Problem Loss of Osseointegration (2408)
Patient Problems Failure of Implant (1924); Osteolysis (2377); Inadequate Osseointegration (2646)
Event Date 01/28/2020
Event Type  Injury  
Event Description
It was reported that, after a left tha had been performed on (b)(6) 2010, the patient experienced lysis around the acetabular cup and loosening of the acetabular cup.A revision surgery was performed on (b)(6) 2020 to treat this adverse event.During this procedure, the acetabular cup was explanted.The patient's health status is unknown.This information was provided by the national joint registry of the united kingdom, as part of a retrospective data collection of patients who underwent either a revision or re-revision surgery where a polarcup shell was part of the hip construct.As such, no further information will be available.
 
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Manufacturer Narrative
H10: a total of 70 revision surgeries involving polarcup devices was reported by the uk national joint registry.104 complaints were opened accordingly.A review of the device history record (dhr), of the complaint history (chr), of the instruction for use (ifu), of the risk management file (rmr) and of past escalation (per) was performed.No relevant supporting clinical information was provided.Based on insufficient information, a thorough clinical assessment cannot be performed at this time.The patient's current condition is unknown and the patient impact beyond the reported events could not be determined.Based on the chr and dhr evaluation performed there is no indication that any failure to meet specification contributed to the reported events.The reported failure modes are outlined in the device instruction for use.The occurrence and severity of the reported issues are in line with the risk plans.The root cause of the reported events was attributed to known inherent risks of the devices.No corrective actions have been initiated.Internal complaint reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP SHELL HA COATED 51 NON CEM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15080096
MDR Text Key296370967
Report Number9613369-2022-00367
Device Sequence Number1
Product Code LPH
UDI-Device Identifier03700421001095
UDI-Public3700421001095
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Study,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number75008719
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/30/2022
Initial Date FDA Received07/21/2022
Supplement Dates Manufacturer Received07/29/2022
Supplement Dates FDA Received08/17/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
PN: 71312372 / LOT: 909963; PN: 75004136 / LOT: 90878
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age79 YR
Patient SexFemale
Patient Weight63 KG
-
-