MAUDE Adverse Event Report: PERFUSION SYSTEMS CANNULA CB67532 VEN RT PVC 32FR CARMEDA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PERFUSION SYSTEMS CANNULA CB67532 VEN RT PVC 32FR CARMEDA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS Back to Search Results
Model Number CB67532
Device Problems Fracture (1260); Air/Gas in Device (4062)
Patient Problems Hypovolemia (2243); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2022
Event Type  Injury  
Event Description
Medtronic received information that during use of a dlp single stage venous cannula, it was reported that about ten minutes after the start of extracorporeal membrane oxygenation (ecmo), a fracture of the cannula appeared while tying the wire (premicron 1) to attach the device to the skin.The wire cut the cannula in two, resulting in massive air entry into the ecmo circuit, since this cannula is positioned on the suction (drainage) part of the circuit.The customer immediately stopped ecmo and clamped the arterial line.No air passage was observed in the arterial part of the circuit neither upstream of the clamp, nor downstream, so no prior gas embolism but it was necessary to redo the cannulation and change the ecmo circuit which resulted in a low-flow of 45-50 mean arterial pressure (map) for 15 minutes and a hemodilution to 7.5g followed by a transfusion of psl (produit sanguin labile - blood product) (2/2/1 minimum).The blood balance was normalized that day.There was no patient impact associated with this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information for b5: medtronic received additional information that the patient did not lose blood, but the pump bubbled ecmo, so a circuit had to be repurposed.The customer used 800 ml of additional hemodilution.The damage was not at the suture site but at the point of attachment of the cannula to the thorax.The cannula was neither heated nor cooled.Correction section d 3 manufacturer name and address: these fields have been updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
The blood balance was normalized that day and no additional adverse patient effects were reported.Correction h6: the fdp/annex a code has been updated to e0305.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Conclusion: complaint is confirmed for damage to the cannula body in the wire wound area.Visual inspection of the returned device found multiple areas of what appears to be damage at the connection end and in the wire wound area of the cannula body.Dimension for wall thickness was taken using calibrated calipers in the area where the wire wound tubing had separated.Each side of the tubing was taken.Dimensions identified on the drawing shows 0.035 +/-.005.Results shows the wall thickness was within on both pieces.It is possible this issue is the result of the user tightening the wire around the wire wound portion of the cannula body.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.Additionally, the instructions for use (ifu) states that these cannula are only indicated for "cardiopulmonary bypass surgery up to six hours or less" and is contraindicated for long-term use.Therefore, these devices are not indicated for use within an extracoporeal membrane oxygen (ecmo) circuit as was the case with the complaint device reported in this product event.Complaints received from june 2021 through fe bruary 2023 for similar model numbers were reviewed and showed no trends warranting escalation related to this occurrence.There were no adverse patient effects as a result of this occurrence.Medtronic has made its supplier aware of this occurrence and will conti nue to monitor for future occurrences and trends.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Device evaluation: visual inspection showed, the device was damaged/split.The reason for return, was confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULA CB67532 VEN RT PVC 32FR CARMEDA
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key15093207
MDR Text Key296472867
Report Number9612164-2022-02801
Device Sequence Number1
Product Code DWF
UDI-Device Identifier00763000192617
UDI-Public00763000192617
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K120988
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 03/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/15/2023
Device Model NumberCB67532
Device Catalogue NumberCB67532
Device Lot Number222045875
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/27/2022
Initial Date FDA Received07/25/2022
Supplement Dates Manufacturer Received07/21/2022
09/06/2022
09/06/2022
01/21/2023
02/24/2023
Supplement Dates FDA Received08/17/2022
09/30/2022
01/04/2023
02/20/2023
03/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
Patient Weight50 KG
-
-