• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/03/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.The customer reported that the damper was broken.As a result, the internal elements in the subject device had become too loose and affected the function of the internal gas compression pump.Confirming the user¿s report, during the evaluation it was discovered that the damper of the manifold unit was indeed broken.Consequently, the manifold unit could not be held firmly in place during operation.The unsecure manifold unit caused the break and deformation of the sockets on the main board.If additional information becomes available following the device evaluation, a supplemental report will be filed.
 
Event Description
The customer returned his loaner-olympus high flow insufflation unit with a complaint of a broken damper.During the device evaluation, it was discovered that the unit had abnormal insufflation due to a damaged channel part.This report is being submitted to capture abnormal insufflation confirmed during the device evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 7 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, it was determined that the internal components were loosened, and internal gas compression pump function was affected because the damper was damaged.The loose manifold unit impacted the inner elements resulting in the breakage and deformation of sockets on the main board (rc board).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura nishigo-mura
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15094079
MDR Text Key304724067
Report Number3002808148-2022-00272
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/03/2022
Initial Date FDA Received07/25/2022
Supplement Dates Manufacturer Received08/04/2022
Supplement Dates FDA Received08/25/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-