Model Number 0998-00-3023-53
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/13/2022
Event Type  malfunction  
Event Description
It was reported that, the cs300 intra-aortic balloon pump (iabp) unit needed a battery replacement.It is unknown under which circumstances the event occurred or if there was any patient involvement.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during a routine check the cs300 intra-aortic balloon pump (iabp) unit needed a battery replacement.There was no patient involvement.
 
Event Description
N/a.
 
Manufacturer Narrative
Updated fields: b4, d9, e1 (site country), g3, g6, h2, h3, h4, h6 (type of investigation, investigation findings, component codes and investigation conclusions), h10 additional customer contact phone: (b)(6).Getinge field service engineer (fse) found replaced the batteries(146-00-0039) on this unit because they expired this month.Unit passed all functional and safety tests performed.Unit was returned to customer and cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15098988
MDR Text Key304291085
Report Number2249723-2022-01865
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/13/2022
Initial Date FDA Received07/25/2022
Supplement Dates Manufacturer Received07/28/2022
02/27/2024
Supplement Dates FDA Received08/22/2022
02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/28/2011
Is the Device Single Use? No
Type of Device Usage A
Patient Sequence Number1
-
-