VENTEC LIFE SYSTEMS, INC VOCSN; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number PRT-00490-001
Device Problem Defective Alarm (1014)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2022
Event Type  malfunction  
Event Description
It was reported that the device needed service.Upon evaluation of the device, ventec observed that it did not provide an o2 concentration alarm as expected during testing.There was no patient involvement associated with the reported event.
 
Manufacturer Narrative
The device was evaluated by ventec where the reported issue of the device not providing an o2 concentration alarm as expected during testing was confirmed.Ventec updated the device's software to version and confirmed that the o2 concentration alarm triggered as expected during the final test.Proper device operation was observed through functional and performance testing.The cause of the device failing to alarm for o2 concentration during the initial device evaluation could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOCSN
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer (Section G)
VENTEC LIFE SYSTEMS, INC
22002 26th ave se
bothell WA 98021
Manufacturer Contact
atul sinha
22002 26th ave se
bothell, WA 98021
2067612204
MDR Report Key15108970
MDR Text Key302762621
Report Number3013095415-2022-00657
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162877
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPRT-00490-001
Device Catalogue NumberPRT-00490-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2022
Initial Date Manufacturer Received 07/07/2022
Initial Date FDA Received07/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/20/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-