• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR GUIDE WIRE ACC.KIT W/COPILOT®; ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR GUIDE WIRE ACC.KIT W/COPILOT®; ACCESSORIES Back to Search Results
Model Number 1003330
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/03/2022
Event Type  Injury  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It is possible that the clamp seal was not entirely closed; thus resulting in the reported leak; however as the device was not returned for analysis this cannot be confirmed.The investigation determined a conclusive cause for the reported difficulties cannot be determined.The treatment appears to be related to the operational context of the procedure as the y-connector had to be held throughout the procedure in order to stop the leakage.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event Description
It was reported that during the procedure, there was leakage in the y-connector of the copilot.The y-connector had to be held throughout the procedure in order to stop the leakage.There were no adverse patient effects and there was no reported clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUIDE WIRE ACC.KIT W/COPILOT®
Type of Device
ACCESSORIES
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 2024168
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15117124
MDR Text Key296753440
Report Number2024168-2022-08259
Device Sequence Number1
Product Code DTL
UDI-Device Identifier08717648013980
UDI-Public08717648013980
Combination Product (y/n)N
Reporter Country CodeTC
PMA/PMN Number
K991102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2024
Device Model Number1003330
Device Catalogue Number1003330
Device Lot Number60359636
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/07/2022
Initial Date FDA Received07/27/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-