• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO. LTD. A4 ANESTHESIA MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO. LTD. A4 ANESTHESIA MACHINE Back to Search Results
Model Number 0630F-PA00001
Device Problem Gas Output Problem (1266)
Patient Problem Tachycardia (2095)
Event Type  Injury  
Manufacturer Narrative
An evaluation of the anesthesia machine was performed and the o2 cell cable was found to be loose causing a leak.The o2 cell cable was tightened and the system was tested and performed as designed.
 
Event Description
It was reported that the a4 anesthesia system was not working while in use on a patient, resulting in the patient rebreathing co2.The patient experienced a period of tachycardia but then returned to normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
A4 ANESTHESIA MACHINE
Type of Device
A4 ANESTHESIA MACHINE
Manufacturer (Section D)
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO. LTD.
1203 nanhuan ave
guangming district
shenzhen, guangming 51810 6
CH  518106
MDR Report Key15118146
MDR Text Key296751961
Report Number2221819-2022-00014
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/27/2022,06/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0630F-PA00001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/27/2022
Event Location Hospital
Date Report to Manufacturer07/27/2022
Initial Date Manufacturer Received Not provided
Initial Date FDA Received07/27/2022
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-