MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 UNK HIP FEMORAL HEAD METAL ASR; METAL FEMORAL HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 UNK HIP FEMORAL HEAD METAL ASR; METAL FEMORAL HEAD Back to Search Results
Catalog Number UNK HIP FEMORAL HEAD METAL ASR
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Ossification (1428); Foreign Body Reaction (1868); Pain (1994); Loss of Range of Motion (2032); Renal Failure (2041); Anxiety (2328); Distress (2329); Discomfort (2330); Metal Related Pathology (4530); Unspecified Tissue Injury (4559)
Event Date 03/31/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Asr litigation record received.Patient alleges pain, discomfort, elevated metal ion, permanent injuries, emotional distress, disability, disfigurement,and patient's ability to perform normal and enjoy regular activities has been impaired.Doi: (b)(6) 2009 - dor: (b)(6) 2021 (right hip).
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
In addition to previous allegation pfs alleges limited mobility, adl, anxiety and metallosis.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
 
Manufacturer Narrative
Product complaint (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Medical records received: after a review of the medical records, the patient was revised to address elevated metal ions, discomfort, chronic kidney disease, hip and back pain, dysfunction, and limited mobility.Operative note reported some heterotopic ossification, upon opening the joint there was a black liquid sludge came out with typical sign of metallosis and wear in the hip.There was some small amount of notching in the femoral neck in the superolateral aspect.Asr cup, head and sleeve were revised.Re-implanted with 66 multi-hole redapt modular cup (competitor), 52 polyethylene ball from the smith & nephew competitor and 1.5 titanium sleeved 28 diameter biolox delta ceramic (depuy).Summit stem was unrevised.There was no further information provided post-revision or any complication.Right hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK HIP FEMORAL HEAD METAL ASR
Type of Device
METAL FEMORAL HEAD
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15121179
MDR Text Key296767655
Report Number1818910-2022-14302
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD METAL ASR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/13/2022
Initial Date FDA Received07/28/2022
Supplement Dates Manufacturer Received08/02/2022
01/23/2023
06/21/2023
04/24/2024
Supplement Dates FDA Received08/03/2022
01/25/2023
06/21/2023
04/29/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL SLEEVE ASR; UNK HIP FEMORAL STEM SUMMIT; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age86 YR
Patient SexMale
Patient Weight102 KG
-
-