Model Number 989803194541 |
Device Problem
Fire (1245)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/04/2022 |
Event Type
malfunction
|
Event Description
|
A philips field service engineer (fse) reported that 8 philips batteries were damaged in a basement fire.These 8 batteries were part of a set of 10, 2 were returned to (b)(4) for investigation, 8 were left in fse¿s basement.There was no patient involvement and not in clinical use.
|
|
Manufacturer Narrative
|
Philips was able to obtain the fire investigator report which stated that "there was nothing seen among the debris that revealed a competent ignition source.The possibility of a problem with any of the rechargeable batteries is something that i cannot eliminate at this time." two batteries not in the fire were evaluated , both were working as designed an no problem could be found.
|
|