U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 105; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 105; GENERATOR Back to Search Results
Model Number 105
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Convulsion, Clonic (2222); Discomfort (2330); Numbness (2415)
Event Date 04/11/2020
Event Type  Injury  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
Patient was referred for explant surgery.It was later noted the patient wanted vns therapy removed as it bothered her (affecting her left arm) and caused numbness.Patient continued to have daily seizures, which she attributed to stress caused by vns therapy.No known surgery has occurred to date.No additional relevant information has been received.
 
Manufacturer Narrative
Date received by manufacturer (mo/day/yr), corrected data: initial report inadvertently listed the wrong date, shouldve been 07/01/2022.
 
Event Description
Additional information received noting that the cause of the increased seizures is low battery.A battery life calculation was performed with the available data and assumptions were made for any gaps found if present.The result revealed 0 yrs remaining until eri = yes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 105
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
dana sprague
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key15133371
MDR Text Key296909893
Report Number1644487-2022-00916
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750054
UDI-Public05425025750054
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/12/2016
Device Model Number105
Device Lot Number203191
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 07/05/2022
Initial Date FDA Received07/29/2022
Supplement Dates Manufacturer Received08/05/2022
Supplement Dates FDA Received08/30/2022
Date Device Manufactured12/11/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age38 YR
Patient SexFemale
-
-