MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OBTRYX II SYSTEM - HALO; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OBTRYX II SYSTEM - HALO; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR Back to Search Results
Model Number M0068505110
Device Problems Detachment of Device or Device Component (2907); Material Integrity Problem (2978); Material Twisted/Bent (2981); Positioning Problem (3009)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/27/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an obtryx ii system - halo device was used during a transobturator sling placement procedure performed on (b)(6) 2022 for the treatment of urinary incontinence.During the procedure, the physician had difficulty in tensioning the sling and the mesh curled.Photos of the device taken during the procedure showed the mesh was stretched and curled.The procedure was then cancelled and rescheduled on (b)(6) 2022.When the patient returned, a piece of plastic sleeve from the mesh assembly was found in the patient and was retrieved by hand.The procedure was completed with another obtryx ii system - halo.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OBTRYX II SYSTEM - HALO
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR STRESS URINARY INCONTINENCE, RETR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15135003
MDR Text Key304356467
Report Number3005099803-2022-04252
Device Sequence Number1
Product Code OTN
UDI-Device Identifier08714729837565
UDI-Public08714729837565
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121754
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM0068505110
Device Catalogue Number850-511
Device Lot Number0027868629
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/07/2022
Initial Date FDA Received07/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age47 YR
Patient SexFemale
Patient Weight90 KG
-
-