Model Number M00522610 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the device analysis, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a resolution clip device was used in the stomach during an endoscopic submucosal dissection (esd) procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip could not detach from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Event Description
|
It was reported to boston scientific corporation that a resolution clip device was used in the stomach during an endoscopic submucosal dissection (esd) procedure performed on (b)(6) 2022.During the procedure, the clip was able to grasp and lock onto tissue; however, the clip could not detach from the catheter to deploy.The procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
Block h6: medical device code a15 captures the reportable event of clip could not detach from catheter.Block h10: investigation results the returned resolution clip device was analyzed, and a visual evaluation noted that the device was returned without the clip assembly and with evidence of full deployment.The device was returned without the over-sheath, only the blue grip was returned attached to the device.Microscopic examination was performed and it was found that the bushing had hit marks.Additionally, the bushing was deformed.Dimensional analysis was performed on the bushing outer diameter, and it was found to be within specification.A dimensional analysis was also performed between the hooks of the bushing, and one side of the hooks was found to be out of specification.No other problems with the device were noted.The reported event of clip could not detach from catheter was not confirmed.Investigation found that one side of the hooks was out of specification, this condition likely happened during or after the procedure, since this condition would have prevent the clip to be assembled in a first instance during the manufacturing process.It is possible that operational factors such as the technique used during removal of the clip, or the interaction between the yoke and the capsule, in order to deploy the clip could have caused the found problems of bushing deformed and hit marks on the bushing.However, the device was returned without the clip assembly and it is important to mention that the presence of this component is very important to the investigation in order to analyze any failure that could contribute with the problem faced by the physician.The returned device review showed no evidence of either the alleged(s) or any defect which could have contributed to the event.Therefore, the most probable root cause for this complaint is no problem detected.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.
|
|