• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. ULTRASONIC PROBE Back to Search Results
Model Number UM-S20-17S
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/03/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, the reported issue was confirmed and there was no other abnormality.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported that blood contamination was noted at the tip of the ultrasonic probe.The event occurred during a postoperative reprocess, prior to a diagnostic endoscopic ultrasonography operation.The same device was used to complete the operation and there was no patient harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record (dhr) found no deviations that could have caused or contributed to the contamination (blood)/leakage of the ultrasonic medium.The dhr confirmed that the subject device was shipped in accordance with the specifications.A definitive root cause for this issue was not established.However, it is probable that the blood entered the device from a perforation at the tip of the sheath.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) which states the following: ¿chapter 4 operation 4.2 observation of the ultrasound image ¿ never push or pull the ultrasonic probe with excessive force or withdraw it into the bending section of the endoscope during probe rotation (real-time mode).Manipulate the ultrasonic probe slowly and carefully when the endoscope is sharply angled or the forceps elevator is raised.Forcefully pushing or pulling the ultrasonic probe may damage it.¿ do not attempt probe rotation while the ultrasonic probe is advanced abruptly.This could result in image flow and impeded or cause irregular rotation.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15138414
MDR Text Key304802341
Report Number3002808148-2022-00511
Device Sequence Number1
Product Code ITX
UDI-Device Identifier04953170368479
UDI-Public04953170368479
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/04/2022
Initial Date FDA Received08/01/2022
Supplement Dates Manufacturer Received09/01/2022
Supplement Dates FDA Received09/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/12/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-