A review of the device history record confirmed that the device was manufactured and tested according to relevant procedures, and shipped according to manufacturer's specifications.After further discussion with the surgeon and op sales rep, it was determined that the kite angle measurement tool was used incorrectly during the surgical procedure.Specifically, there was confusion of how to interpret the markings on the tabs with avatar and the "l" and "r" on each of the two tabs, respectively.It was interpreted that the "r" meant a right curve, and since it was a lenke 1 right scoliosis, this is where the arrow of the alignment tool was fixed.(for clarity, the "r" and the avatar indicate that the mid-c system is place on the right side of the patient and not that the pathology (the curve) is to the right.The root cause of this confusion was a lack of understanding the training.The company conducted additional training to clarify this issue with the surgeon and op sales rep on the day after the index procedure which was the day before the revision procedure.The revision result was reported to be excellent.The company's incident rate of extender misalignment is 2.19% · the risk of "extender and mid-c system not properly aligned" has been assessed and found to be acceptable (b)(4).
|