Model Number URF-P6
Device Problem Degraded (1153)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
During inspection and testing, service found the opening of the forceps stopper was scraped due to physical stress which resulted in a leak.In addition, angulation was insufficient due to elongation of the angle wires, there were scratches on multiple parts of the device, and the image guide bundle was broken due to physical stress.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented accordingly.
 
Event Description
An olympus employee reported there was an air leak from the channel of the loaner device.The subject device was sent to an olympus service center for evaluation.During inspection and testing, service found the forceps stopper was scraped.This report is being submitted for the malfunction found during evaluation (scraping in the channel port).There was no effect on the patient due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that stress of repeated use, external factors, or handling caused the event.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15155212
MDR Text Key298028528
Report Number9610595-2022-00566
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340802
UDI-Public04953170340802
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/07/2022
Initial Date FDA Received08/03/2022
Supplement Dates Manufacturer Received08/31/2022
Supplement Dates FDA Received09/19/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-