RESPIRONICS, INC. DREAMSTATION BIPAP AUTOSV; VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING Back to Search Results
Model Number DSX900H11C
Device Problem Degraded (1153)
Patient Problem Pulmonary Hypertension (4460)
Event Date 09/28/2021
Event Type  Injury  
Event Description
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.The patient alleged having pulmonary hypertension.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a continuous positive airway pressure (cpap) device's sound abatement foam.The patient alleged to experience pulmonary hypertension.There was no medical intervention required by the patient.The reported event of pulmonary hypertension and its reported severity was reviewed by the manufacture's clinical expert.This event is assessed as not related to the device in this case.Based on the available information, the manufacture concludes no further action is necessary.There was no medical intervention required by the patient.The device has not yet returned to the manufacturer for evaluation.The patient has swapped out the deice at the distributor and at this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.Section(s) b1, b2, has changed related to the complaint changing from the reported adverse event to a product problem.Section h1 has changed to reflect a malfunction.Section h6 health effect- impact code, type of investigation findings and investigation conclusions has been updated.
 
Manufacturer Narrative
The manufacturer previously submitted mdr 2518422-2022-67943-1 with incorrect sections b1, b2, h1, h6.Corrections to previous mdr are made in this report as follows.Section b1 was corrected to adverse event and product problem.(only product problem was checked in previous mdr) section b2 was corrected to other serious or important medical events.(previously it was blank) section h1 was changed from malfunction to serious injury.Section h6- health impact code was updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION BIPAP AUTOSV
Type of Device
VENTILATOR, CONTINUOUS, NON-LIFE-SUPPORTING
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key15164916
MDR Text Key297246300
Report Number2518422-2022-67943
Device Sequence Number1
Product Code MNS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX900H11C
Device Catalogue NumberDSX900H11C
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 09/28/2021
Initial Date FDA Received08/04/2022
Supplement Dates Manufacturer Received02/09/2023
02/09/2023
Supplement Dates FDA Received03/23/2023
06/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/20/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
Patient Outcome(s) Other;
-
-