• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US UNKNOWN ANKLE IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US UNKNOWN ANKLE IMPLANT Back to Search Results
Catalog Number UNK ANKLE
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Pain (1994); Joint Laxity (4526); Unspecified Tissue Injury (4559); Swelling/ Edema (4577)
Event Date 03/03/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This complaint is from a literature source.The following literature cite has been reviewed: benson a, alvine f, alvine g.Sixteen-year follow-up of talar body prosthesis and total ankle arthroplasty after traumatic talar extrusion: a case report.Jbjs case connect.2021 mar 3;11(1).Doi: 10.2106/jbjs.Cc.20.00543.Pmid: 33657023.Objective and methods: (b)(6) patient received a custom agility total ankle replacement 2 tears after accidental excision of her natural talus in 2002.The patient received a talar component, tibial component, talar stem, and tibial insert in addition to a tibial-fibular syndesmosis fused with allograft and two unknown screws.In 2009, the patient experienced pain and swelling and x-rays revealed a collapsed tibial component rotated into valgus alignment.No treatment was provided as the pain resolved.In 2019, the patient again reported pain and joint instability.A ct scan revealed subluxation and impingement secondary to wear of the tibial insert.Upon entering the joint, metallosis due to articulation on the talar component on the tibial component secondary to tibial insert wear was identified and debrided.There was some navicular bone erosion noted.The tibial and talar components were well-fixed, with some wear, and retained.The migration of the tibial component and joint laxity was confirmed.The patient received a revision of the poly insert.The procedure was completed without complications.Lot, model and catalog number are not available, but the suspected depuy device possibly associated with reported adverse events: agility ankle: tibial component, tibial insert, talar component, talar stem.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary
=
> no device associated with this report was received for examination.This complaint was opened to document complaints derived through a journal article review.Follow-ups were done to try and obtain additional information from the author of the journal article.No further information was received.Device history lot
=
> a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN ANKLE IMPLANT
Type of Device
ANKLE IMPLANT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15174911
MDR Text Key297332918
Report Number1818910-2022-15203
Device Sequence Number1
Product Code HSN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Literature,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK ANKLE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/23/2022
Initial Date FDA Received08/05/2022
Supplement Dates Manufacturer Received08/26/2022
Supplement Dates FDA Received08/29/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-