MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VISERA ELITE XENON LIGHT SOURCE Back to Search Results
Model Number CLV-S190
Device Problem Device Emits Odor (1425)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/11/2022
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer¿s allegation was not confirmed.The evaluation uncovered the lock release was due to detection lever wear.There were no other abnormalities found.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available this report will be supplemented.
 
Event Description
During an unknown procedure, the visera elite xenon light source generated a sound, and a burnt smell was detected without an error message being displayed.The procedure was completed using a similar device.There were no reports of patient harm associated with this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.It has been over 10 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, a root cause could not be identified.The phenomenon was not reproduced.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The cause of the reported event cannot be conclusively determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA ELITE XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer Contact
masaharu hirose
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8-061
JA   961-8061
426422891
MDR Report Key15176477
MDR Text Key304963824
Report Number3002808148-2022-00731
Device Sequence Number1
Product Code NWB
UDI-Device Identifier04953170301995
UDI-Public04953170301995
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-S190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/11/2022
Initial Date FDA Received08/05/2022
Supplement Dates Manufacturer Received09/14/2022
Supplement Dates FDA Received10/12/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse