• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS RUBELLA IGG; ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS RUBELLA IGG; ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA Back to Search Results
Model Number RUBELLA IGG
Device Problem High Test Results (2457)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2022
Event Type  malfunction  
Event Description
There was an allegation of a questionable rubella igg elecsys result from cobas 8000 - cobas e 602 module serial number 16p2-08.The rubella igg elecsys result was 19.22 iu/ml (positive).The result by rub igg (siemens) was 1.30 iu/ml (negative).The questionable result was not reported outside of the laboratory.
 
Manufacturer Narrative
The investigation was ongoing.
 
Manufacturer Narrative
The reagent expiration date was provided as october 2022.The calibration and qc data were inconspicuous.The customer was not using the required rack adapters for 13 mm diameter tubes when the event occurred.The investigation did not identify a product problem.The cause of the event could not be determined as no sample material was available for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS RUBELLA IGG
Type of Device
ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key15176578
MDR Text Key305248382
Report Number1823260-2022-02301
Device Sequence Number1
Product Code LFX
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K072617
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRUBELLA IGG
Device Catalogue Number04618793190
Device Lot Number58677700
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/11/2022
Initial Date FDA Received08/05/2022
Supplement Dates Manufacturer Received09/21/2022
Supplement Dates FDA Received10/17/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient SexFemale
-
-