Model Number 16-02-85
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/14/2022
Event Type  malfunction  
Event Description
Livanova received report that a heater cooler system 3t gave ee error during maintenance.Reportedly, the device worked afterwards but the customer requested the machine to be checked.No additional information is available up to date and the event is conservatively considered reportable.There was no patient involvement.
 
Manufacturer Narrative
There was no patient involvement.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in rock hill, south carolina.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
H.10: a livanova field service representative was dispatched to the facility to investigate the device and could not confirm the reported issue.The device was found to be working within specifications and no deviation could be found.Subsequent functional verification testing was completed without further issues and the unit was returned to service.Through follow-up communication livanova learned that the customer noted the error code while filling the device tanks with water.Based on service activity findings and collected information, the reportability decision has been changed to not reportable event.Indeed, the error is most likely one of the three (3) error codes which are expected to be displayed on the device during device filling phase and to disappear once filling is completed.Therefore, the issue is not likely to result in serious injury and it is always detectable prior device use.
 
Event Description
See initial report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key15182703
MDR Text Key297900225
Report Number9611109-2022-00400
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817900528
UDI-Public010403381790052811170515
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/14/2022
Initial Date FDA Received08/08/2022
Supplement Dates Manufacturer Received09/13/2022
Supplement Dates FDA Received10/12/2022
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/15/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-