Model Number 0998-00-0800-65 |
Device Problem
Circuit Failure (1089)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/21/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during pre delivery inspection, the cardiosave intra-aortic balloon pump (iabp) when the unit was turned on the confirmation sound was generated but no sound was heard.The issue was found with the backplane board and it was replaced, once replaced the issue was resolved.There was no patient involvement.
|
|
Manufacturer Narrative
|
A supplemental report will be submitted upon completion of our investigation.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) stated pre-delivery inspection is carried out at the kobe office.When he pressed the main power switch, a confirmation sound was generated but no sound was heard.Fse replaced of the backplane board (because there is a speaker that produces sound on the backplane board).After replacement, a confirmation sound was generated and there was no problem.
|
|