Model Number FG540000 |
Device Problem
Fire (1245)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/01/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).The hardware investigation has begun but it has not been completed at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Event Description
|
It was reported that an unknown patient underwent an unknown ablation procedure with a carto® 3 system.There was a burning smell in the power cord region.Burning smell in the system cart power cord region.No patient consequences were reported.Fire hazard is mdr-reportable.
|
|
Manufacturer Narrative
|
On 8-aug-2022, bwi received additional information regarding the event.The issue was not solved.The medical team is waiting for the appraisal of the equipment division.The root may be malfunction of the electrical converter.They have the report yet regarding if root cause was related to carto; it¿s the responsibility of the equipment division.No errors or messages observed.They did not see any smoke, just the smell.Actins taken were to turn off the equipment.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Manufacturer Narrative
|
On 15-sep-2022, the product investigation was completed.It was reported that an unknown patient underwent an unknown ablation procedure with a carto® 3 system.There was a burning smell in the power cord region.Device evaluation details: field service engineer (fse) arrived at the hospital and checked the system.The fse found that cart system did not present any defects.There was no smell of burning anywhere.The fse performed tests with clinical engineering and head nursing.The system is operational.A manufacturing record evaluation was performed for the carto 3 system # 14549, and no internal actions related to the reported complaint condition were identified.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|