DEPUY INTERNATIONAL LTD - 8010379 ADAPTER SLEEVES 12/14 +8; ASR HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Catalog Number 999800318
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Cyst(s) (1800); Dyspnea (1816); Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Inflammation (1932); Necrosis (1971); Pain (1994); Anxiety (2328); Distress (2329); Deformity/ Disfigurement (2360); Fibrosis (3167); Metal Related Pathology (4530); Muscle/Tendon Damage (4532); Unspecified Tissue Injury (4559); Physical Asymmetry (4573); Swelling/ Edema (4577)
Event Date 06/08/2022
Event Type  Injury  
Event Description
Litigation complaint received ad 7 jul 2022.Claim letter alleges breathing difficulties, very high levels of metal toxicity.Patient was revised was due to metallosis, a tumor, cysts and necrotizing tissue.Doi: 18 years ago, dor: (b)(6) 2022, unk hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Occupation: lawyer.(b)(4).Investigation summary: no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
The patient was alleged to experience pain, increased metal levels in blood including cobalt and chromium, injuries, emotional distress, and disfigurement.Doi: (b)(6) 2007.Dor: (b)(6) 2022.Right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
 
Event Description
Medical records were received and stated the following: pfs alleges injuries on tendons, ligaments, tissues, bones within right hip.Plaintiff suffered extreme pain, trochanteric fractured, elevated metal ions, pseudotumor, metallosis and anxiety.After review of the medical records the patient was revised to address hip pain, elevated metal ions, fractured greater trochanter, pseudotumor, metallosis, trunnionosis and vertical cup.Operative note reported metallosis in the posterior capsule.There was fluid in the hip, trunnionosis in the femoral head.Pseudotumor appear to track up the iliopsoas.There was leg length discrepancy.Doi: (b)(6) 2007;dor: (b)(6) 2022; right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: h6 (clinical codes) h6 clinical code: - swelling/ edema (e2338) is used to capture swelling and edema.- unspecified tissue injury (e2015) is used to capture soft tissue injury and bone injury.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTER SLEEVES 12/14 +8
Type of Device
ASR HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key15198440
MDR Text Key297590425
Report Number1818910-2022-15517
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040627
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 08/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number999800318
Device Lot Number2372485
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 07/25/2022
Initial Date FDA Received08/10/2022
Supplement Dates Manufacturer Received01/03/2023
03/08/2023
07/06/2023
Supplement Dates FDA Received01/05/2023
03/08/2023
07/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/19/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
ADAPTER SLEEVES 12/14 +8.; ASR ACETABULAR CUPS 56.; ASR UNI FEMORAL IMPL SIZE 49.; ASR UNI FEMORAL IMPL SIZE 49.; SUMMIT DUOFIX TAP SZ5 HI OFF.; UNK HIP ACETABULAR CUP ASR.; UNK HIP FEMORAL SLEEVE ASR.; UNKNOWN HIP FEMORAL STEM.
Patient Outcome(s) Required Intervention;
Patient Age58 YR
Patient SexMale
Patient Weight107 KG
-
-